Ductless Fume Hood F-DLFH103 is a floor standing unit with microprocessor control system, and LED display. It comes with 1380 × 730 × 745 mm chamber dimension, 520 mm maximum opening and 9 levels adjustable air speed. Equipped with UV lamp for effective sterilization of working environment, centrifugal blower and electronic control system. Designed with three side transparent toughened glass window, anti-UV motorized sliding design front window and waterproof sockets, 4 caster wheel for easy movement, water sink, water tap, gas tap, foot switch, exhaust, active carbon filter and chemical resistant phenolic resin work table.
Internal dimension (W × D × H) | 1380 × 730 × 745 mm |
External dimension (W × D × H) | 1500 × 840 × 2150 mm |
Work surface height | 750 mm |
Maximum opening | 520 mm |
Air velocity | 0.3 to 0.8 m/s |
Noise | ≤ 68 dB |
LED lamp | 16 W × 1 |
UV lamp | Emission of 253.7 nm for most efficient decontamination |
Blower | Built-in centrifugal blower, speed adjustable with 9 levels |
Front window | Motorized 5 mm anti-UV toughened glass |
Material | |
Exterior | Cold-rolled steel with anti-bacteria powder coating. |
Work table | Chemical resistant phenolic resin |
Waterproof sockets | 2 pcs |
Power | 500 W |
Power supply | AC 220 V ± 10%, 50/60 Hz; 110 V ± 10%, 60 Hz |
Packaging dimension (W × D × H) | 1650 × 1120 × 1680 mm |
Gross weight | 290 kg |
Ductless type of fume hood |
9 levels adjustable air speed |
Resistance to weak acid and alkali |
UV lamp for effective sterilization of working environment |
Built-in centrifugal blower |
Three side transparent glass window, anti-UV motorized sliding design front window |
Microprocessor control system |
Digital LED display |
4 Casters for easy movement |
Chemical resistant phenolic resin work table |
Ergonomically design 8 degree front slope |
External material made up of cold rolled steel with anti-bacterial powder coating |
Recovery function from power failure |
Alarm when filter work time is 3500 hours |